A. Anastasi, D. Babusci, F. Baffigi, G. Cantatore, D. Cauz, G. Corrada, S. Dabagov, G. Di Sciascio, R. Di Stefano, C. Ferrari, A.T. Fienberg, A. Fioretti, L. Fulgentini, C. Gabbanini, L.A. Gizzi, D. Hampai, D.W. Hertzog, M. Iacovacci, h, M. Karuza, k, J. Kaspar, P. Koester, L. Labatb, S. Mastroianni, D. Moricciani, G. Pauletta, L. Santi, G. Venanzoni

Nuclear Instruments and Methods in Physics Research A788, 43–48 (2015)

Abstract

The new muon (g−2) experiment E989 at Fermilab will be equipped with a laser calibration system for all the 1296 channels of the calorimeters. An integrating sphere and an alternative system based on an engineered diffuser have been considered as possible light distributors for the experiment. We present here a detailed comparison of the two based on temporal response, spatial uniformity, transmittance and time stability.