[free_download_btn]
V. De Leo, G. Claps, F. Cordella, G. Cristoforetti, L. A. Gizzi, P. Koester, D. Pacella and A. Tamburrino
Abstract
We present an innovative X-ray spectroscopy system to address the complex study of the X-ray emissions arising from laser–target interactions, where the emissions occur within extremely brief intervals from femtoseconds to nanoseconds. Our system combines a Gas Electron Multiplier (GEM) detector with a silicon-based Timepix3 (TPX3) detector. These detectors work in tandem, allowing for a spectroscopic radiation analysis along the same line of sight. With an active area of 10 x 10 cm2, the GEM detector allows for 1D measurements for X-ray energies (2–50 keV) by utilizing the full 10 cm gas depth. The high-energy part of the radiation beam exits through a downstream side window of the GEM without being absorbed in the gas volume. Positioned side-on at the GEM detector’s exit, the TPX3 detector, equipped with a pixelated sensor (55 μm x 55 μm; active area 14 mm x 14 mm), uses its full 14 mm silicon sensor to detect hard X-rays (50–500 keV) and gamma rays (0.5–10 MeV). We demonstrate the correct operation of the entire detection system and provide a
detailed description of the Timepix3 detector’s calibratio
- Versione
- Scarica 2
- Dimensioni file 5.58 MB
- Conteggio file 1
- Data di Pubblicazione Novembre 17, 2023
- Ultimo aggiornamento Novembre 17, 2023